Synchronisation of Ultrasound Scanners, Synchronization of External Equipment
Ultrasound research may require the synchronization of ultrasound machines and other equipment. Telemed developed optional modules to equip ArtUs, SmartUs, MicrUs, and ClarUs beamformers with such feature. Such optional modules (I/O modules) are available for such modifications (see notes below) and should be ordered separately:
Signals / scanners
ArtUs
SmartUs
MicrUs
ClarUs
- Ultrasound Line output
yes
yes
yes
yes
- Ultrasound Frame output
yes
yes
yes
yes
- Ultrasound Line input
yes
no
no
no
- Ultrasound Frame input
yes
no
no
no
- ScanStart output
yes
yes
yes
yes
- ScanStart input
yes
yes
no
no
Signals description:
- Ultrasound Line output - at the falling edge system starts acquiring ultrasound data from tissue for one ultrasound line
- Ultrasound Frame output - at the falling edge system starts acquiring ultrasound data from tissue for one ultrasound frame
- Ultrasound Line input - system starts acquiring ultrasound data from tissue for one ultrasound line approximately in 20 us after the rising edge of the signal
- Ultrasound Frame input - system starts acquiring ultrasound data from tissue for one ultrasound frame approximately in 20 us after rising edge of the signal
- ScanStart output - system generates logic level "0" in Freeze mode and logic level "1" in scan mode
- ScanStart input - system starts scan (from new frame) when logic level "1" and stops scan when logic level "0"
Example: MicrUs EXT-1H with installed I/O module (Ultrasound Line & Ultrasound Frame outputs).